Revision:1.0(2023-01-12)
281
2. Use a crosshead screwdriver to remove the four M3x6 countersunk head screws to take off the
cover of the battery management board.
1 Four M3X6 countersunk head screws
2 Cover of the power management board
3. Use a crosshead screwdriver to remove the four M3x8 Phillips pan-head screws to take off the
two connection terminals, and then use a crosshead screwdriver to remove the eight M3x8
Phillips pan-head screws to take off the power management board.
1 Two connection terminals
2 Four M3x8 Phillips pan-head screws
3 Eight M3x8 Phillips pan-head screws
7.49.3 Commissioning and Verification
None
Diagnostic Ultrasound System
Service Manual
7 FRU Replacement