250
FRU Code FRU Name Lower-Level FRU
Code
Lower-Level FRU
Name
115-083596-00 Gas spring assembly
(FRU)
7.31.2 Disassembly and Assembly
Precautions
Protect the exterior surface of the parts.
Tools Required
Art knife, crosshead screwdriver (M4), flathead screwdriver, socket head wrench (M5), and
circlips
Fixtures Required
None
Specific Steps
1. Refer to the disassembly procedures in 7.2.2 Disassembly and Assembly, the section of
removing the main unit box, 7.22.2
Disassembly and Assembly, 7.23.2 Disassembly and
Assembly
, 7.30.2 Disassembly and Assembly, 7.27.2 Disassembly and Assembly, and 7.28.2
Disassembly and Assembly, and take off each assembly.
2. Refer to the disassembly procedures in 7.47.2
Disassembly and Assembly, 7.29.2 Disassembly
and Assembly
, 7.38.2 Disassembly and Assembly, 7.37.2 Disassembly and Assembly, and 7.46.2
Disassembly and Assembly, and take off each assembly. Then, remove the back board of the
battery box.
3. Use a crosshead screwdriver to remove the 12 M4X12 crosshead combination screws that
connect the cover, and take off the cover.
Diagnostic Ultrasound System
Service Manual
7 FRU Replacement

Table of Contents

Other manuals for Mindray TEX20

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray TEX20 and is the answer not in the manual?

Mindray TEX20 Specifications

General IconGeneral
BrandMindray
ModelTEX20
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals