Revision:1.0(2023-01-12)
278
7.48 Battery Pack indicator board (FRU)
7.48.1 General Information
FRU Code FRU Name Lower-Level FRU
Code
Lower-Level FRU
Name
115-083602-00 Battery pack indicator
board (FRU)
Function description:
It main displays the battery capacity status and the AC indicator status.
7.48.2 Disassembly and Assembly
Precautions
Protect the exterior surface of the parts.
Tools Required
Crosshead screwdriver (M3)
Fixtures Required
None
Specific Steps
1. Refer to the disassembly procedures in 7.47.2 Disassembly and Assembly and 7.50.2
Disassembly and Assembly to remove the battery box cover and battery.
2. Use a crosshead screwdriver to remove the two M3x6 countersunk head screws to take off the
pressure plate of the indicator board (with the sheet metal part).
Diagnostic Ultrasound System
Service Manual
7 FRU Replacement

Table of Contents