209
7.9 Probe Board Assembly (FRU)
7.9.1 General Information
FRU Code
FRU Name Lower-Level FRU
Code
Lower-Level FRU
Name
115-083578-00 Probe Board
Assembly (FRU)
Function description:
It connects and expands the probe channel with the transmitting and receiving channel in the
main unit box.
7.9.2 Disassembly and Assembly
Precautions
None
Tools Required
Anti-static gloves
Crosshead screwdriver (M3)
Fixtures Required
None
Specific Steps
1. Refer to the disassembly procedure in 7.20.2 Disassembly and Assembly to take off the main
core.
Diagnostic Ultrasound System
Service Manual
7 FRU Replacement

Table of Contents