Revision:1.0(2023-01-12)
242
7.26 Wireless Probe Charging Socket PCBA
(FRU)
7.26.1 General Information
FRU Code
FRU Name Lower-Level FRU
Code
Lower-Level FRU
Name
115-083594-00 Wireless Probe
Charging Socket
PCBA
7.26.2 Disassembly and Assembly
Precautions
Protect the exterior surface of the parts.
Tools Required
Art knife, crosshead screwdriver (M3\M4), flatheaded screwdriver, and socket head wrench (M5)
Fixtures Required
None
Specific Steps
1. Refer to the disassembly procedures in 7.2.2 Disassembly and Assembly, the section of
removing the main unit box, 7.22.2
Disassembly and Assembly, 7.23.2 Disassembly and
Assembly
, 7.30.2 Disassembly and Assembly, and 7.27.2 Disassembly and Assembly, and take
off each assembly.
2. Remove the connection between the charging station cable socket and the Spring Cable
socket. Then, use a crosshead screwdriver to remove the two M3×8 crosshead combination
Diagnostic Ultrasound System
Service Manual
7 FRU Replacement

Table of Contents

Other manuals for Mindray TEX20

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray TEX20 and is the answer not in the manual?

Mindray TEX20 Specifications

General IconGeneral
BrandMindray
ModelTEX20
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals