Revision:1.0(2023-01-12)
254
FRU Code FRU Name Lower-Level FRU
Code
Lower-Level FRU
Name
115-083611-00 Retractable AC Cable
Assembly (America)
Function description:
It is used to organize the power cord of the trolley.
7.33.2 Disassembly and Assembly
Precautions
Protect the exterior surface of the parts.
Tools Required
Art knife, crosshead screwdriver (M3\M4), flatheaded screwdriver, and socket head wrench (M5)
Fixtures Required
None
Specific Steps
1. Refer to the disassembly procedures in 7.2.2 Disassembly and Assembly, the section of
removing the main unit box, and 7.47.2
Disassembly and Assembly to take off each assembly.
Take off the battery box assembly and disconnect the Retractable AC Cable Assembly from the
power cable and grounding cable of the Mobile Trolley Connecting Board.
2. Lay down the trolley, use a crosshead screwdriver to remove the eight M4x12 crosshead
combination screws that connect the Base Bottom Cover, so as to take off the Retractable AC
Cable Assembly.
Diagnostic Ultrasound System
Service Manual
7 FRU Replacement

Table of Contents

Other manuals for Mindray TEX20

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray TEX20 and is the answer not in the manual?

Mindray TEX20 Specifications

General IconGeneral
BrandMindray
ModelTEX20
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals