7.27.3 Commissioning and Verification
None
7.28 Back Housing for Upper Column
7.28.1 General Information
FRU Code FRU Name Lower-Level FRU
Code
Lower-Level FRU
Name
043-013036-00 Back Housing for
Upper Column
7.28.2 Disassembly and Assembly
Precautions
After assembling or disassembling the Back Housing for Upper Column, lift the trolley to the
highest position. Protect the surface of each part during assembly and disassembly.
Tools Required
Art knife and crosshead screwdriver (M4)
Fixtures Required
None
Specific Steps
1. Lift the trolley to the highest position.
2. Use an art knife to remove the three screw hole identification stickers on the rear of the Back
Housing for Upper Column. Then use a crosshead screwdriver to remove the three M4X12
Diagnostic Ultrasound System