7.13.2 Disassembly and Assembly
Precautions
None
Tools Required
Anti-static gloves
Crosshead screwdriver (M3)
Fixtures Required
None
Specific Steps
1. Refer to the disassembly procedures in 7.7.2 Disassembly and Assembly, 7.4.2 Disassembly and
Assembly
, 7.8.2 Disassembly and Assembly, 7.20.2 Disassembly and Assembly, and 7.9.2
Disassembly and Assembly to take off the main unit box assembly.
2. Remove the 13 Phillips pan-head screws on the cover of the main board box, and then take off
the cover of the main board box.
1 Cover of the main board box
2 13 Phillips pan-head screws
Diagnostic Ultrasound System
Service Manual
7 FRU Replacement