258
7.35 Base Bottom Cover
7.35.1 General Information
FRU Code FRU Name Lower-Level FRU
Code
Lower-Level FRU
Name
043-010610-00 Base Bottom Cover
7.35.2 Disassembly and Assembly
Precautions
Protect the exterior surface of the parts.
Tools Required
Art knife, crosshead screwdriver (M3\M4), flatheaded screwdriver, and socket head wrench (M5)
Fixtures Required
None
Specific Steps
1. Refer to the disassembly procedures in 7.2.2 Disassembly and Assembly, the section of
removing the main unit box, and 7.47.2
Disassembly and Assembly to take off the monitor
assembly, main unit box, and battery box cover. Then remove the fixing screws from the battery
box assembly and take off the battery box assembly. Refer to the disassembly procedures in
7.38.2
Disassembly and Assembly, 7.37.2 Disassembly and Assembly, and 7.46.2 Disassembly
and Assembly
to take off each assembly.
Diagnostic Ultrasound System
Service Manual
7 FRU Replacement

Table of Contents

Other manuals for Mindray TEX20

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray TEX20 and is the answer not in the manual?

Mindray TEX20 Specifications

General IconGeneral
BrandMindray
ModelTEX20
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals