Revision:1.0(2023-01-12)
40
No. Name Description
3. Power button/indicator Used to power up or down the system. The indicator is
on after the system is powered up.
4. Power output port Supplies power to optional peripheral devices.
The auxiliary power output outlet in the system is
used to supply power to the recommended
peripheral devices. Do not connect other devices
to the outlet, otherwise, the rated output power
may be exceeded and failure may be resulted.
5. Equipotential terminal
Used for equipotential connection, that balances the
protective earth potentials between the system and other
electrical equipment.
6. Caster Used for securing or moving the system
7.
Retractable AC power
cord
AC power cord.
8. Wireless charging module Used with the wireless charging module to charge the
battery for the main unit.
9. Batteries Used to supply power when there is no AC power supply.
10. Storage Basket
Used to store sundries such as medical records and
paper towels.
11. Probe port Socket connecting the probe and the main unit.
12. Touch screen and monitor Provides operator-system interface or control.
13. Lifting adjustment key Pressed to adjust the height of the main unit support.
14. Handle Used to lift up and down or move.
15. Printer fixing bracket Used to fix the printer.
2.1.3 Peripherals Supported
Type Model
Digital graph/text printer CANON TS708
Black and white video printer
MITSUBISHI P95DW-N
Black and white video analog printer SONY UP-X898MD
Diagnostic Ultrasound System
Service Manual
2 Product Knowledge