• Do not immerse main unit into liquid.
• When disinfecting the main unit, do not disinfect the metal parts and interfaces on the
main unit to prevent corrosion.
• Do not pour liquid on the main unit or let liquid enter the shell.
• If you accidentally pour the liquid on the main unit, please disconnect the power
immediately, wipe the liquid, and contact the maintenance personnel or Mindray.
• Never use abrasive materials (such as steel wool or silver polish), or erosive cleaners
(such as acetone or acetone-based cleaners) for cleaning.
• Please dilute and use the cleaner and disinfectant according to the disinfectant
manufacturer's instructions.
• After cleaning and disinfecting, please check the machine carefully. If you find aging
or damage, stop using it immediately.
• Follow local regulations when selecting and using the recommended disinfectant.
• The disinfectant recommended in this section is only used to disinfect the main unit,
and cannot be used to disinfect other parts such as probes.
• Wear clinical gloves to prevent infection.
Recommended Disinfectants
The following table lists the main unit disinfectants recommended by Mindray.
Reagent name Reagent name Type
75% medicinal alcohol Common disinfectant Solutions
99.5% isopropanol Common disinfectant Solutions
Hydrogen peroxide Common disinfectant Solutions
0.5% sodium hypochlorite Common disinfectant Solutions
SONOTM ULTRASOUND
WIPES
Advanced Ultrasound Solutions Inc. Wet wipes
Schulke mikrozid® Sensitive
Wipes
Schulke Wet wipes
Mikrobac Tissues Bode Chemie Gmbh Wet wipes
Dianerkang ® surface
disinfectant
Dianerkang Solutions
Dianerkang ® disinfectant
spray
Dianerkang Solutions
Protex Disfectant Spray Parker Laboratoires Inc. Spray
Diagnostic Ultrasound System
Service Manual
5 Maintenance