284
Protect the exterior surface of the parts.
Tools Required
Crosshead screwdriver (M3)
Fixtures Required
None
Specific Steps
1. Refer to the disassembly procedure in 7.47.2 Disassembly and Assembly to take off the Battery
Box cover.
2. Use a crosshead screwdriver to remove the two Phillips pan-head screws on one side of the
battery box and take off the pressure plate.
1 Pressure plate
3. Remove the cable terminals, open the battery door on this side for about 90°, and lift up the
battery door, so as to take off the Battery Box Fan Assy(FRU).