Revision:1.0(2023-01-12)
83
1 Lower part of the main unit box assembly
2 Push back
3 Spring plunger
4 Pull down and reset
After installation, ensure that the spring plunger correctly rebounds.
The method of determining whether the spring plunger correctly rebounds is as
follows:
1. Pull the spring plunger and release it. The spring plunger can correctly rebound.
2. Without pulling the spring plunger, you cannot lift the main unit box assembly with
your hands.
If the spring plunger fails to correctly rebound, the main unit box assembly may fall
unexpectedly, damaging the device and hurting people.
Diagnostic Ultrasound System
Service Manual
3 Installation