DMTA-20073-01EN, Rev. C, November 2015
Preparing the RollerFORM Liquid Chamber
39
3. Preparing the RollerFORM Liquid Chamber
Prior to performing inspections, you must fill the RollerFORM scanners acoustic
wheel liquid chamber with noncorrosive glycol or water, and you need to make sure
that it is devoid of any air bubbles.
3.1 Filling the Acoustic Wheel’s Liquid Chamber
To be able to perform data acquisition with the RollerFORM scanner, you must first
fill the acoustic wheel’s liquid chamber with the appropriate liquid, using the
supplied manual pump.
3.1.1 Choosing a Liquid to Fill the Chamber
Depending on the environmental and/or safety constraints of the inspection to be
carried out, you can use noncorrosive glycol, distilled or deionized water, or ordinary
tap water to fill the liquid chamber of the RollerFORM scanner.
The following list outlines the conditions under which you should consider using
each of these liquids:
Glycol, noncorrosive
Olympus recommends using Propylene Glycol Dynalene PG, or a noncorrosive glycol
with the same specifications.
Using noncorrosive glycol inside the liquid chamber provides the following
benefits:
NOTE

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Olympus RollerFORM and is the answer not in the manual?

Olympus RollerFORM Specifications

General IconGeneral
BrandOlympus
ModelRollerFORM
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals