Maintenance
91
To leave the liquid chamber to dry
1. Remove the acoustic wheel from the RollerFORM.
For more information, see “Removing the Acoustic Wheel from the RollerFORM
Scanner” on page 71.
2. Unscrew the OUT valve completely, and then remove it (see Figure 6-17 on
page 91).
Store the valve safely to find it easily when putting it back in place.
Figure 6-17 The location of the OUT valve
3. Gently and repeatedly squeeze the tire, until all the water is drained out (see
Figure 6-18 on page 92).