9.3.6.2 Number of logged points (3.6.2)
Sets number of data points to capture.
Setting 10 to 100000
Default 50 000
Maximum number of logged points per the following table:
Number of PV inputs
Maximum points
8 10 002
7 13 336
6 16 670
5 20 004
4 23 338
3 33 340
2 50 010
1 100 020
9.3.6.3 Logging interval (3.6.3)
Se
ts interval in seconds between log entries.
Setting 10 to 600
Default 600
9.3.6.4 Logging behavior (3.6.4)
De󹪝nes beha
vior when log is full.
Setting Overwrite oldest
Fill and stop
Default Overwrite oldest
Parameter assignment
9.3Maintenance and diagnostics (3)
SITRANS LR500 series with mA/HART
140 Operating Instructions, 03/2024, A5E51099898-AA