6.3.3.2 Instructions speci󹪝c to hazardous area installations (Reference ATEX Directive
2014/34/EU, Annex II, 1.0.6 and UK Regulations SI 2016/1107))
The following instructions apply to equipment covered by certi󹪝cate number CSANe
23ATEX1113X, CSANe 23ATEX1114X, CSAE 23UKEX1087X and CSAE 23UKEX1088X.
1. For use and assembly, refer to the main instructions.
2. The equipment is certi󹪝ed for use as Category 1GD equipment per CSANe 23ATEX1113X and
CSAE 23UKEX1087X; Category 1D equipment per CSANe 23ATEX1114X and CSAE
23UKEX1088X.
3. As Category 1GD equipment per CSANe 23ATEX1113X and CSAE 23UKEX1087X, the
equipment may be used with 󹪞ammable gases and vapors with apparatus group IIC, IIB and
IIA or with 󹪞ammable dusts with apparatus group IIIC, IIIB, IIIA with the temperature classes
as speci󹪝ed per Temperature. (Page169)
4. As Category 1D equipment per CSANe 23ATEX1114X and CSAE 23UKEX1088X, the
equipment has a degree of ingress protection of IP66/68 and may be used with 󹪞ammable
dusts with apparatus group IIIC, IIIB, IIIA with the temperatures as speci󹪝ed per Temperature
(Page169).
5. The equipment has not been assessed as a safety related device (as referred to by Directive
2014/34/EU, clause 1.5 and UK Regulation SI 2016/1107).
6. Installation and inspection of this equipment shall be carried out by suitably trained
personnel in accordance with the applicable code of practice (EN 60079-14 and EN
60079-17 in Europe and the UK).
7. The equipment is non-repairable.
8. The certi󹪝cate numbers have an ‘X’ su󹪟x, which indicates that special conditions for safe use
apply. Those installing or inspecting this equipment must have access to the certi󹪝cates.
9. If the equipment is likely to come into contact with aggressive substances, then it is the
responsibility of the user to take suitable precautions that prevent it from being adversely
a󹪜ected, thus ensuring that the type of protection is not compromised.
Aggressive substances: for example, acidic liquids or gases that may attack metals, or
solvents that may a󹪜ect polymeric materials.
Suitable precautions: for example, establishing from the material’s data sheet that it is
resistant to speci󹪝c chemicals.
Connecting
6.3Hazardous area installations
SITRANS LR500 series with mA/HART
60 Operating Instructions, 03/2024, A5E51099898-AA

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens SITRANS LR580 and is the answer not in the manual?

Siemens SITRANS LR580 Specifications

General IconGeneral
BrandSiemens
ModelSITRANS LR580
CategoryTransmitter
LanguageEnglish

Related product manuals