102
4. Turn the dial of the Control Knob to select the option you want.
5. Press the Control Knob to confirm and accept the settings.
To disable the Pulse Oximetry function from the Main screen select and press
Menu – Advance Setting – SpO2 enabled – Disable.
For information regarding the Pulse Oximetry screen operation see page 120.
PNEUMATIC NEBULIZER
A Nebulizer is a pneumatic device that uses compressed gas to deliver
aerosolized medication that can be inhaled by patients. During nebulization, the
flow is synchronized with the inspiratory phase of each breath. Note that the
Nebulizer works during the inspiratory phase only.
For proper installation of the Nebulizer see page 46.
CAUTION The nebulizer is operational for all tidal volumes greater than 200ml. Do not
activate the nebulizer if the set tidal volume is below 200ml.
To set the Nebulizer:
1. From the Main screen select Menu – Advanced Settings.
2. In the Advanced Settings screen select Nebulizer. The current setting is
shown in the right column of the screen (Figure 91).
Figure 91: Nebulizer Selected
3. Press the Control Knob. The Nebulizer pop-up window appears. The default
time is OFF (Figure 92)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE iVent 201 Series and is the answer not in the manual?

GE iVent 201 Series Specifications

General IconGeneral
BrandGE
ModeliVent 201 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals