1 Introduction
29
MRI CONDITIONAL TESTS SPECIFICATIONS
The iVentâ„¢201 MR conditional has been performed to demonstrate that there are
no known hazards in a specified MR environment with specified conditions.
The specified MR environment was:
Model:
Magnetom Trio, Siemens Medical Solutions, Erlangen, Germany, 3
Tesla, horizontal static magnetic field, active shielded
Image frequency:
123.192952 MHz
Gradient System:
max. 40 mT/m, Slew rate: 200 mT/m/ms
Software:
Numaris/4, syngo MR 2004A 4VA25A
Coil:
Body coil for transmitting & receiving signals
Field Strength:
3 Tesla
Spatial gradient:
0.12 mT/m
dB/dt
200mT/m/ms
RF:
13-231 Watts
SAR:
0.2 – 92.2
Spin echo:
SE 2D
Time of echo (TE):
20 ms
Time of repetition
(TR):
1000 ms
Field of view (FoV):
400 x 400mm
Matrix
256 x 256 pixel
FoV phase:
100%
Flip angle
90°
Bandwidth/ Pixel:
100 Hz
Slice thickness:
5 mm
No. of slices:
1
Slice orientation:
Coronal
No. of averages:
1
Phase encoding:
Right –left (RL)
iVentâ„¢ 201
configuration during the test:
Ventilation Type:
SIMV Volume Control
Tidal Volume:
500mL
PEEP:
5 cmH20
FiO2:
50
Power source:
AC and battery
Distance from
isocenter:
100 Gauss (approximately 3 meters from the isocenter, this field
strength must be measured with a gauss meter)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE iVent 201 Series and is the answer not in the manual?

GE iVent 201 Series Specifications

General IconGeneral
BrandGE
ModeliVent 201 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals