174
CLEANING THE VENTILATOR
Clean the exterior of the ventilator when necessary with a soft damp cloth
moistened with one of the following cleaning solutions:
ï‚· Water
ï‚· Mild detergent or soapy water
Do not allow liquid to penetrate into the ventilator.
PLANNED MAINTENANCE
Observing the following maintenance intervals helps keep the iVent
TM
201
operating trouble-free. Follow the iVent
TM
201 Service Manual procedure for
instructions.
Table 13: Planned Maintenance Procedures
Frequency Part Maintenance
As necessary Ventilator enclosure
Wipe clean the
exterior
Patient tubing Replace and
perform O.V.T.
Cooling air filter and
vents
Clean and replace
air filter as needed
O2 sensor Replace the O2
sensor
Every 500 hours or 1
month or whichever
comes first
Air Inlet Filter Remove the used air
inlet filter and
replace it with a new
inlet filter
1500 hours or 3
months of use
O2 System O
2
Calibration
Battery
Charge the battery
Ventilator Perform VVT
3,000 hours or every 6
months of use
(whichever comes
first)
O2 System O2 calibration
Battery Deep discharge and
recharge
Ventilator Perform VVT
Annually Air Inlet Filter
Replace Air Inlet
Filter
Inlet muffler
assembly.
Outlet muffler
assembly
Replace Inlet muffler
assembly and Outlet
muffler assembly