Appendix G: Accessories
245
3. Connect the external exhalation valve control tube (blue tube, C in Figure
177) to its port.
4. Before use on a patient, perform the “Operation Verification Test (O.V.T)” as
described on page 127.
Instructions for Use With a Pneumatic Nebulizer
NOTE To use a pneumatic nebulizer with the ventilator, connect an external high-
pressure oxygen supply to the ventilator, and select the high-pressure oxygen
supply option from the Advanced Setting menu as described on page 94.
1. Connect the nebulizer tube to the nebulizer outlet on the front panel of the
ventilator (G in Figure 177).
2. Install the nebulizer with the desired medication to the patient circuit after
the Wye connector and the flow sensor.
3. Connect the free end of the nebulizer tube to the respective nebulizer port.
4. For instructions on activating the Nebulizer see page 102.
REUSEABLE PATIENT BREATHING CIRCUIT
These instructions refer to the following reusable circuits:
Breathing Circuit REF
Patient Circuit for adult M/U 620B0008-02/ M1161145
Breathing Circuit W. Trap M/U 620B0023-02/ M1162028
Breathing Circuit Heated W M/U 620B0024-02/ M1162029
Contents
Non-sterile, reusable patient circuit
Exhalation Valve with Exhalation Control Tube
Flow Sensor with 2 Channel Tubing
22 mm Cuffed Breathing Tube
2 caps for Operational Verification Test (O.V.T)
Warnings and Cautions
WARNING US Federal laws restrict this device for sale or use by
or on the order of a physician.
The patient circuit must at all times be used with the
One-Way valve inserted. Failure to insert the One-
Way Valve could result in insufficient oxygenation of
the patient and inadequate ventilation of the