1 Introduction
17
O2 SUPPLY SPECIFICATIONS
Parameter
High Pressure Pressure Range 40 –75psi (2.8-5.1 bar).
Warning: The exact oxygen pressure which the
iVent
TM
201 uses depends on the version you
have. Check the back label (shown in Figure 11)
for the pressure levels of your unit.
Low Pressure Maximum flow 15L/min or 0.5 psi
VENTILATION PERFORMANCE AND CONTROLLED
PARAMETERS
Parameter
Respiratory Rate 1 – 12 ±1, 12 – 80 ± 2 bpm
Tidal Volume 50-2000 mL
Accuracy of Tidal Volume
Delivery
±10% or ±10 mL whichever is greater
Accuracy of Exhale Tidal
volume Measurement
±15% above 100 mL from actual reading
or ±10 cc below 100 mL
Inspiratory Pressure Limit 5 to 80 ±5 cmH
2
O
Inspiratory Time Adaptive Time
TM
or 0.2 to 3 ±10% sec
Peak Flow (PIF) Adaptive Flow
TM
or to 120 ±10% L/min
Spontaneous to 180 ±10% L/min
Oxygen Mix (FiO
2
) 21% to 100% ±10% of the set value or ±3%
absolute, whichever is greater
PEEP 0 to 40 ±1 cmH
2
O or ±10%, whichever is
greater
Trigger Sensitivity
1 to 20 L/min Flow, Off
- 0.5 to – 20 cmH
2
O Pressure, Off
PSV 0 to 60 ±10% cmH
2
O
Positive Pressure Relief
Valve
80 cmH
2
O
Pneumatic Nebulizer 5 to 240 minutes
Controlled Pressure 5 to 80 cmH
2
O ±5 cmH
2
O
FiO
2
at Power up 21%, 40%, 60%, 100% (configurable)
Configurable Humidifier
Setting (Purging Cycle)
HME (10 minute), Heated (1 minutes), Off