220
SUMMARY
The classification of breaths, for the purpose of this manual, are summarized in
the following table.
Table 15: Summary of Breath Types
Breath Type
Initiation Controlled
parameters
Limiting
parameters
Termination
Mandatory Breath
Ventilator
Volume Control
By Ventilator
(time trigg.)
Flow Pressure or
Time
By Ventilator
Set V
T
is
delivered
Assist Volume
Control
By Patient
(pt. trigg.)
Flow Pressure or
Time
By Ventilator
Set V
T
is
delivered
Ventilator
Pressure
Control
By Ventilator
(time trigg.)
Pressure
Level
Volume By Ventilator
Inspiratory
Time is
elapsed
Assist Pressure
Control
By Patient
(pt. trigg.)
Pressure
Level
Volume By Ventilator
Inspiratory
time is
elapsed
Manual By Operator Volume or
Pressure
Level
Pressure,
Volume, or
Time
By Ventilator
Set V
T
is
delivered or
Inspiratory
time is
elapsed
Breath Type Initiation Controlled
parameters
Limiting
parameters
Termination
Patient Breath
Spontaneous
Breath
By Patient
(pt. trigg.)
Baseline
Pressure
(PEEP)
Inspiratory
Time
By Patient
(flow drop or
pressure
increase)
Pressure
Support Breath
By Patient
(pt. trigg.)
Target
Pressure
(above PEEP)
Inspiratory
Time
By Patient
(flow drop or
pressure
increase)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE iVent 201 Series and is the answer not in the manual?

GE iVent 201 Series Specifications

General IconGeneral
BrandGE
ModeliVent 201 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals