3 Operating the iVentTM201 – Setting Modes and Parameters
65
ï‚· Tidal volume
ï‚· Pressure limit
ï‚· Oxygen level
ï‚· Peak flow
ï‚· Pressure support
ï‚· PEEP
ï‚· Triggers
ï‚· Inspiratory time
ï‚· Rise Time
 Flow Termination – Exhale sensitivity (Essens)
CHANGING VENTILATION PARAMTERS – OVERVIEW
CAUTION Only a fully qualified professional should change these settings
The iVent
TM
201 provides two different ways to change ventilation parameters:
through the Main screen and through the Mode Parameters screen.
NOTE To quickly change an individual parameter setting, the fastest way is
through the Main screen. If you need to adjust several settings at once,
navigate to the Mode Parameters screen and make each required
adjustment. When you have set each parameter, you can accept all of
them at once.
To change ventilation parameters through the Main Screen:
1. Return to the Main screen -- if you are not already there, press the Clear
button until you return to it.
2. Turn the Control Knob to select the parameter you want to adjust.
3. Press the Control Knob.
4. Adjust the setting to the required value.
5. Press the Control Knob again to confirm your changes and return to the Main
screen.
To change ventilation parameters through the Mode Parameter Screen:
1. From the Main window, choose the Mode Selection option and select the
required ventilation mode. The Parameters screen appears.
2. Dial the Control Knob to select the particular setting you wish to change.
3. Press the knob to display the setting’s control window.
4. Turn the knob until the parameter has been changed to the required level or
value.