EasyManuals Logo
Home>GE>Medical Equipment>iVent 201 Series

GE iVent 201 Series User Manual

GE iVent 201 Series
290 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #157 background imageLoading...
Page #157 background image
6 Alarms
145
ï‚· High and Low O2
ï‚· This alarm is updated automatically when setting the FiO2 parameter and
can also be adjusted from the Alarm Settings screen
ï‚· Leak %
NOTE For the Pulse Oximetry related alarms see page 122
The following alarm values are configurable:
ï‚· Volume Limit Reached (for pressure controlled modes only).
This alarm can be adjusted from the Tidal Volume pop-up window in the Main
screen, whereas the other alarms are set from the Alarm Settings screen.
ï‚· Audio Alarm Volume
ï‚· Inverse I:E Ratio Alarm
ï‚· Tidal Volume not delivered
 Patient Circuit Disconnect (Enable/ Disable – On/Off)
The following alarms are always operational:
A/C Power disconnect High PEEP
Low Battery Low O
2
pressure
Battery Empty Check Sensor
Over Temperature Sensor Disconnect
Patient Circuit Failed Auto Start
Service Notice Need calibration
CHANGING INDIVIDUAL ALARM SETTINGS
Select and press Alarm Settings in the Main Menu to view the Alarm Settings
screen (Figure 133):

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE iVent 201 Series and is the answer not in the manual?

GE iVent 201 Series Specifications

General IconGeneral
BrandGE
ModeliVent 201 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals