Setting the Alarm Volume
NOTE The Alarm volume default is level 8.
To set the alarm volume:
1. From the screen select Menu – Alarm Setting - Options.
2. On the Alarm Options screen, select the first option Alarm Volume Level.
3. Press the Control Knob to display the Alarm Volume pop-up window (Figure
143).
Figure 143: The Alarm Volume Pop-Up Window
4. Turn the dial to adjust the alarm volume from 1 (low) to 10 (high). As you turn
the dial, the iVent
TM
201 produces a sample of the level. If you set the level
under 2 or over 8, a Caution! pop up window appears.
5. Press the Control Knob to confirm the settings.
6. You are returned to the Alarm Settings screen. Dial and select Accept to
accept the Alarm Options settings.
Setting the Inverse I:E Ratio Alarm
WARNING The default setting for Inverse I:E Ratio Alarm is ON. Before disabling
any alarms, ensure that you are operating in accordance with facility
policy and patient needs.
To disable or enable the inverse I:E Alarm:
1. From the Main screen select Menu – Alarm Settings - Options. The Alarm
Option menu is displayed.
2. Turn the Control Knob until Inverse I:E Ratio Alarm is selected.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE iVent 201 Series and is the answer not in the manual?

GE iVent 201 Series Specifications

General IconGeneral
BrandGE
ModeliVent 201 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals