7 Care, Maintenance and Tests
181
O2 supply
O2 hose
Starting the VVT
The following procedures explain how to start the Ventilator Verification Test
(VVT).
To start the VVT procedures:
1. Connect a breathing circuit to the iVent
TM
201.
2. Make sure the ventilator is plugged into a working AC supply and switch
on the ventilator’s power.
3. Select a patient weight to enable the ventilator to enter standby mode.
CAUTION Prior to performing the VVT, the unit must be in a “warm” condition.
Connect the breathing circuit to a test lung and operate the ventilator
for at least 15 minutes.
4. Proceed to the Main Menu from the Main screen.
5. From the Main menu, select Maintenance.
6. A caution pop-up window appears. Select Yes to proceed.
7. Dial the Control Knob to select Ventilator Verification Test, and press.
8. The Ventilator Verification pop-up window appears (Figure 158).
Figure 158: The Ventilator Verification Pop-Up Window
9. Select Start and press the Control Knob.
The Alarm Sound Test
The following procedures explain how to perform the Alarm Sound Test.