7 Care, Maintenance and Tests
181
O2 supply
O2 hose
Starting the VVT
The following procedures explain how to start the Ventilator Verification Test
(VVT).
To start the VVT procedures:
1. Connect a breathing circuit to the iVent
TM
201.
2. Make sure the ventilator is plugged into a working AC supply and switch
on the ventilator’s power.
3. Select a patient weight to enable the ventilator to enter standby mode.
CAUTION Prior to performing the VVT, the unit must be in a “warm” condition.
Connect the breathing circuit to a test lung and operate the ventilator
for at least 15 minutes.
4. Proceed to the Main Menu from the Main screen.
5. From the Main menu, select Maintenance.
6. A caution pop-up window appears. Select Yes to proceed.
7. Dial the Control Knob to select Ventilator Verification Test, and press.
8. The Ventilator Verification pop-up window appears (Figure 158).
Figure 158: The Ventilator Verification Pop-Up Window
9. Select Start and press the Control Knob.
The Alarm Sound Test
The following procedures explain how to perform the Alarm Sound Test.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE iVent 201 Series and is the answer not in the manual?

GE iVent 201 Series Specifications

General IconGeneral
BrandGE
ModeliVent 201 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals