106
Figure 96: The Restore Defaults Warning Pop-Up Window
4. Turn the Control Knob to Confirm and press.
5. The Restore Default Params pop-up window appears, and instructs you to
select a Default Weight. Choose a Default Weight by dialing the Control
Knob. The factory default is 70+kg.
6. Press the Control Knob again. The iVent
TM
201’s settings return to their
defaults and the Main screen is displayed.
DEFAULT PARAMETERS SETTINGS
The following tables summarize the default parameters and alarm values for the
iVent
TM
201 by patient weight:
Table 1: Default Start Parameters for Volume Ventilation for 10 kg to 70 kg
Patient Weight bpm Rate Tidal Volume Max. Pressure (PLimit)
(Kg) (lb) (bpm) (mL) (cMH
2
O)
10 22 30
70
18
15 33 25
105
20
20 44 20
140
25
30 66 18
210
28
40 88 16
280
30
50 110 14
350
35
60 132 12
420
35
70+ 154 12
490
35