Measurement and Analysis
364 Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
• Type the correct weeks or days.
• Select OK.
The system makes the following changes:
• GA (LMP) is now GA (GA) and shows the age you entered.
• In the Patient Data section, the GA changes.
• In the Patient Data section, the EDD (LMP) changes to
EDD (GA) and shows an updated date, using the GA you
entered. In addition, the LMP is erased.
• FetusPos - type information about the fetus position.
• PLAC - type information about the placenta.
Note: To return from a graph to the scan display press
Worksheet or Freeze.
Fetal Trending
When you have ultrasound data for more than one exam for a
patient, you can use the data to look at fetal trending on the
Fetal growth curve graphs. Fetal trending requires that a LMP
value is entered in the Patient information screen.
1. Press Worksheet.
2. Press Graphs and select the desired measurement to
display.
3. Press More and then Plot Both.
The system automatically finds the data from previous
ultrasound exams, and displays it on the graph with the
present data.
Figure 7-66: Fetal Trending