Safety
Vivid S5/Vivid S6 User Manual 39
R2424458-100 Rev. 2
Electrical safety
Internally connected peripheral devices
The system, together with peripheral devices, such as video
tape recorders and printers, meets UL 60601-1 and
IEC 60601-1 (1988) standards for electrical isolation and
safety. These standards are applicable only when the specified
peripheral devices are plugged into the AC outlets provided in
the unit.
External Connection of other peripheral
devices
Other external devices, such as laser cameras, printers, VCRs
and external monitors, usually exceed allowable leakage limits
and, when plugged into separate AC outlets that are then
connected to the unit, are in violation of patient safety
standards. Suitable electrical isolation of such external AC
outlets may be required in order to meet UL2601-1 and
IEC 60601-1 (1988) standards for electrical leakage.
CAUTION
External devices can be used only if CE marked and in
compliance with related standards (EN 60601-1 or EN 60950).
Conformance to EN 60601-1-1 (2000) must be verified.
External devices meeting EN 60950 should be kept outside of the
patient environment, as defined in IEC 60601-1-1 (2000).

Table of Contents

Other manuals for GE Vivid S5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid S5 and is the answer not in the manual?

GE Vivid S5 Specifications

General IconGeneral
BrandGE
ModelVivid S5
CategoryMedical Equipment
LanguageEnglish

Related product manuals