Safety
Vivid S5/Vivid S6 User Manual 49
R2424458-100 Rev. 2
Acceptable devices
The Patient Environmental devices shown on the previous
page are specified to be suitable for use within the PATIENT
ENVIRONMENT.
Unapproved devices
Accessories, options, and supplies
CAUTION
DO NOT connect any probes or accessories without approval by
GE within the PATIENT ENVIRONMENT.
See "Peripheral Update for EC countries" on page 46 for more
information.
CAUTION
DO NOT use unapproved devices.
If devices are connected without the approval of GE, the warranty
will be INVALID.
Any device connected to the Vivid S5/Vivid S6 must conform to
one or more of the requirements listed below:
• IEC standard or equivalent standards appropriate to devices.
• The devices shall be connected to PROTECTIVE EARTH
(GROUND).
CAUTION
Unsafe operation or malfunction may result. Use only the
accessories, options and supplies approved or recommended in
these instructions for use.