IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
3-40 Section 3-6 - Configuration
PRELIMINARY
Section 3-7
Connectivity Overview
3-7-1 Physical Connection
There are several possible connection methods, as outlined below.
3-7-2 Stand-alone Vivid™ S60/Vivid™ S70
No network connection needed.
3-7-3 “Sneaker Net” Environment
No network connection needed.
Use removable media to move data from the Vivid™ S60/Vivid™ S70 to another unit.
3-7-4 Wired Ethernet from Vivid™ S60/Vivid™ S70 to a Workstation
Either of these situations may apply:
• Direct Cable Connection from Vivid™ S60/Vivid™ S70 to a workstation via a Crossover Cable.
You will only need a Crossover Cable for network (TCP/IP) use to connect the two units this way.
a.) Connect one end of the crossed network cable to the network connector on the Vivid™ S60/
Vivid™ S70.
b.) Connect the other end to the network connector to the Workstation.
• Connection via a Peer-to-Peer network.
You will need a network hub and one network cable for each unit connected to the hub.
• Connection via Hospital Network.
You will need one network cable to connect the Vivid™ S60/Vivid™ S70 to a wall jack on the
hospital’s network.
3-7-5 Connection from Vivid™ S60/Vivid™ S70 to a DICOM Server on a Network
You will need one network cable.
1) Connect one end of the cable to the Ethernet connector on Vivid™ S60/Vivid™ S70.
2) Connect the other end of the cable to the wall jack.
NOTE: If a Peer-to-Peer network is connected to the hospital’s network, you may connect the Vivid™ S60/
Vivid™ S70 to the Peer-to-Peer network.