IRECTION FR091521, REVISION 1VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
4-2 Section 4-2 - General Procedures
PRELIMINARY
Section 4-2
General Procedures
4-2-1 Overview
Some procedures are used more often than others. The intention of this section is to keep the most used
procedures in one place.
CAUTION
Ultrasound system requires all covers.
Operate this Ultrasound system only when all board covers and
frame panels are securely in place. The covers are required for
safe operation, good Ultrasound system performance and
cooling purposes.
WARNING
WARNING
Energy Control and Power Lockout for Vivid™ S60/Vivid™
S70.
When servicing parts of the Ultrasound system where there is
exposure to voltage greater than 30 volts:
1. Follow LOCK OUT/TAG OUT procedures.
2. Turn off the breaker.
3. Unplug the Ultrasound system.
4. Maintain control of the Ultrasound system power plug.
5. Wait for at least 30 seconds for capacitors to discharge as
there are no test points to verify isolation.
6. Remove/disconnect the battery, if present.
Ultrasound System components may be energized.

Table of Contents