IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-157
PRELIMINARY
7.) Disconnect the up-down/swivel cable from the handle to be replaced and slide the handle sideways
to remove it from the supporting shaft - Figure 8-192.
8-6-7-5 Swivel and Up-Down Handle Installation Procedure
1.) Install the appropriate handle on the support shaft.
2.) Return spring/s and securing clip/s to secure the handle/s.
3.) Return the protective metal plate and secure it with the retaining screw (previously removed).
4) Install the Probe Shelf as described in the Probe Shelf Installation Procedure on page 8 - 166.
5.) Install the Operator Panel as described in the Operator Panel Keyboard Assembly Installation
Procedure on page 8 - 47.
6) Lower the console to the maximum down position.
7.) Turn ON power to the system.
Figure 8-192 Release Cable and Slide Handle Sideways