IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-59
PRELIMINARY
8-3-13 Operator Panel Keycaps Replacement Procedure
8-3-13-1 Tools
Use the appropriate flat screwdriver, as indicated in the Keycap replacement procedure.
8-3-13-2 Time Required
5 min
8-3-13-3 Preparation
None
8-3-13-4 Operator Panel Keycap Removal Procedure
1.) Carefully insert the end of a small flat screwdriver in the gap between the Keycap you wish to
remove and the surface of the Operator Panel.
2.) Gently lever off the defective Keycap, until it is completely loosened from its base.
3.) Remove the defective Keycap.
8-3-13-5 Operator Panel KeyCap Installation Procedure
1.) Carefully place the appropriate new Keycap in position on the Operator Panel, taking care to place
the plastic alignment pin in the correct position so that the Keycap is the right way up and reads
correctly.
2.) Gently push the Keycap down until it snaps into position.
FRU Part #
Refer to Table 9-11 on page 9-10.
Figure 8-68 Removing Keycap from Operator Panel
Insert end of
Screwdriver
in gap Here

Table of Contents

Other manuals for GE Vivid S60N

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid S60N and is the answer not in the manual?

GE Vivid S60N Specifications

General IconGeneral
BrandGE
ModelVivid S60N
CategoryMedical Equipment
LanguageEnglish

Related product manuals