IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
8-58 Section 8-3 - Control Console Components - Replacement Procedures
PRELIMINARY
8-3-12 Rotary Knob Replacement Procedure
NOTE: The following instructions are applicable to replacement of Rotary Knobs on either on the Operator
Panel, or on the Touch Screen. The Knobs Kit contains Rotary Knobs for both.
8-3-12-1 Tools
None.
8-3-12-2 Time Required
5 min
8-3-12-3 Preparation
None
8-3-12-4 Rotary Knob Removal Procedure
1.) Grip the defective knob between your finger and thumb
2.) Lift straight up and remove from the control shaft - see Figure 8-67 below.
8-3-12-5 Rotary Knob Installation Procedure
1.) Carefully place a new Rotary Knob in position on the Operator Panel (or Touch Screen), taking care
to place the plastic alignment pin in the correct location.
2.) Gently push the Rotary Knob down until it snaps into position.
FRU Part #
Refer to Table 9-11 on page 9-10.
Figure 8-67 Removing a Rotary Knob

Table of Contents

Other manuals for GE Vivid S60N

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid S60N and is the answer not in the manual?

GE Vivid S60N Specifications

General IconGeneral
BrandGE
ModelVivid S60N
CategoryMedical Equipment
LanguageEnglish

Related product manuals