IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
Chapter 3 - System Setup 3-51
PRELIMINARY
Section 3-9
Options Setup
3-9-1 Software Options
Most of the options for Vivid™ S60/Vivid™ S70 are activated by installing a password (alphanumeric
text string).
• For installation instructions, see: Software Options Configuration on page 3 - 39.
3-9-2 Color Video Printer Setup
NOTE: The Color Video Printer is an option.
An external Color Video Printer may be connected to one of the USB ports on Vivid™ S60/Vivid™ S70.
NOTE: SONY UPD-2X-MD is a medical device so it can be placed near the Vivid™ S60/Vivid™ S70. The USB
cable is 5 meters long, so the printer must be placed within reach of this cable.
Follow these steps to connect the printer to the Vivid™ S60/Vivid™ S70:
1) Connect the USB cable to the printer and to one of the USB ports on Vivid™ S60/Vivid™ S70.
2) Select the correct Power cable and connect it to the printer and to the mains power outlet.
3) Switch on the power switch on the printer.
NOTE: To be able to use the printer, the printer must be selected on the Vivid™ S60/Vivid™ S70.
3-9-3 USB Flash Card Setup
There is no special setup procedure for use of a USB Flash Card.
3-9-4 Wireless Network Adaptor
The Wireless Network adaptor should be connected to the lower USB socket on the Vivid S60N/Vivid
S70N Rear I/O Peripheral/Accessory Connector Panel.
NOTE: It is possible to connect the Wireless Network adaptor either once the system is powered ON, or after
shutdown.
All software drivers for the Wireless Network are pre-installed for the designated Wireless Network
adaptor only.
WHEN CONNECTING PERIPHERALS TO THE VIVID™ S60/VIVID™ S70 SYSTEM,
UNLESS THE PERIPHERAL IS MEDICAL-GRADE EQUIPMENT IT SHOULD BE
CONNECTED TO THE ISOLATED USB CONNECTOR.
IMPORTANT During Stand-by mode, it is NOT recommended to introduce or remove USB
devices; this may cause the system to lock-up during the boot-up procedure.
DO NOT ATTEMPT TO USE A DIFFERENT TYPE OF WIRELESS NETWORK ADAPTOR.
THE ULTRASOUND SYSTEM IS AN EXTREMELY SENSITIVE AND COMPLEX
MEDICAL SYSTEM. ANY UNAUTHORIZED PERIPHERALS MAY CAUSE SYSTEM
FAILURE OR DAMAGE!