IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-27
PRELIMINARY
8-2-12 Rear Folder Box Replacement Procedure
8-2-12-1 Tools
8-2-12-2 Time Required
5 min
8-2-12-3 Preparations
Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in Power Shut Down on page 4 - 7.
8-2-12-4 Rear Folder Box Removal Procedure
1) Remove all contents from the folder box.
2) Working from the rear of the system, hold the rear folder box with both hands and carefully lift it
upwards to release the two plastic hooks from the two supporting pins (located one each side of the
Support Column - illustrated in Figure 8-26).
3.) Remove the rear folder box.
FRU Part #
Refer to Table 9-9 on page 9-8.
Figure 8-26 Releasing the Rear Folder Box from the Support Column
Supporting Pin
Plastic Hook