IRECTION FR091521, REVISION 1VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
4-34 Section 4-3 - Functional Checks
PRELIMINARY
4-3-8 Probe/Connectors Check
NOTE: Probes can be connected at any time, whether the unit is ON or OFF
Related information:
Electrical Safety Tests on page 10 - 15
CAUTION
Take the following precautions with the probe cables:
Keep away from the wheels.
Do not bend.
Do not cross cables between probes.
Table 4-2 Probe and Connectors Checks
Step Task Expected Result(s)
1
Press Probe on the Operator Panel. A list of the connected probes will pop
up on the screen.
2
If not already selected, use the trackball to select the
desired probe.
An application menu for the desired
probe is listed on the screen.
3
Trackball to the desired application.
• Press Select to launch the application.
To change application without changing the current
probe, press Appl. on the Operator Panel.
The selected application starts.
4
Verify no missing channels. All channels are functioning.
5
Verify there's no EMI/RFI or artifacts specific to the
probe.
No EMI/RFI or artifacts.
6
Check the probe in each active connector slot. It will display pictorial data each time.
7
Do a leakage test on the probe. It passes the test.
8
Repeat this procedure for all available probes.

Table of Contents

Other manuals for GE Vivid S60N

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid S60N and is the answer not in the manual?

GE Vivid S60N Specifications

General IconGeneral
BrandGE
ModelVivid S60N
CategoryMedical Equipment
LanguageEnglish

Related product manuals