IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
8-36 Section 8-3 - Control Console Components - Replacement Procedures
PRELIMINARY
8-3-3 Touch Screen Rear Cover Replacement Procedure
8-3-3-1 Tools
Phillips screwdriver.
8-3-3-2 Time Required
15 min
8-3-3-3 Preparation
Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in Power Shut Down on page 4 - 7.
8-3-3-4 Touch Screen Rear Cover Removal Procedure
Note: The Touch Screen Rear Cover is fastened to the back of the Touch Screen by 3 screws
through securing tabs located on the base of the cover and 2 screws at the top - Figure 8-34.
1) Check the console is raised to the maximum up position.
Make sure it is aligned in the central position (not pulled to one side or the other).
2) Remove the Touch Screen Service Cover:
3.) Remove the 3 securing screws from the base of the cover - Figure 8-37.
FRU Part #
Refer to Table 9-9 on page 9-8.
Figure 8-36 Touch Screen Rear Cover - Securing Tabs and Screws
Touch Screen Service Cover Removal Procedure on page 8 - 34
Figure 8-37 Removing the Rear Cover (Base) Securing Screws
Internal ViewExternal View

Table of Contents

Other manuals for GE Vivid S60N

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid S60N and is the answer not in the manual?

GE Vivid S60N Specifications

General IconGeneral
BrandGE
ModelVivid S60N
CategoryMedical Equipment
LanguageEnglish

Related product manuals