D
IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-51
PRELIMINARY
8-3-10 Alphanumeric Keyboard Replacement Procedure
NOTE: The Alphanumeric Keyboard is supplied as an integral part of the OPIO Bottom Cover.
In the event that a Vivid S60/Vivid S70 ultrasound scanner that previously did not have the
Alphanumeric Keyboard option fitted and now requires replacement of the OPIO bottom cover, it is
necessary to order and fit the Alphanumeric Keyboard complete with OPIO bottom cover (supplied as
one FRU part).
8-3-10-1 Tools
Phillips screwdriver.
8-3-10-2 Time Required
30 min
8-3-10-3 Preparation
Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in Power Shut Down on page 4 - 7.
8-3-10-4 Alphanumeric Keyboard Removal Procedure
1.) Remove the Operator Panel Keyboard Assembly:.
2.) Carefully place the Operator Panel Keyboard Assembly upside-down on a flat, clean, stable surface.
Figure 8-57 Alphanumeric Keyboard/OPIO Bottom Cover Assembly
When performing these procedures, take precautions to avoid damage of
electrostatic-sensitive components. Always have the ESD wrist strap
connected either to the DIB chassis or to the GND plug at the rear of the
scanner, and to your hand.
If a battery is present, first remove the battery as it contains stored energy.
Refer to Battery Removal Procedure on page 8 - 145.
Refer to Table 9-12 on page 9-11.
• Operator Panel Keyboard Assembly Removal Procedure on page 8 - 44