EasyManuals Logo
Home>GE>Medical Equipment>Vivid S60N

GE Vivid S60N User Manual

GE Vivid S60N
478 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #243 background imageLoading...
Page #243 background image
D
IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-13
PRELIMINARY
8-2-5 DVD and Printer Cover Replacement Procedure
8-2-5-1 Tools
Phillips screwdriver.
8-2-5-2 Time Required
5 min
8-2-5-3 Preparations
Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in Power Shut Down on page 4 - 7.
8-2-5-4 DVD and Printer Cover Removal Procedure
Note: The DVD and Printer cover is secured to the mechanical platform by 4 snap-lock securing
clips on the inner side of the cover.
1) Remove the right side cover:
2.) Grip the bottom of the cover and pull it towards you to release it from the snap-lock securing clips
(Figure 8-30). Remove the cover.
8-2-5-5 DVD and Printer Cover Installation Procedure
1) Return the cover to the right of the system, carefully aligning the 4 snap-lock securing clips with the
securing pins.
2.) Push the cover until it clicks in position.
3.) Install the right side cover.
FRU Part #
Refer to Table 9-8 on page 9-7.
Right Side Cover Removal Procedure
Figure 8-7 Removing the DVD and Printer Cover
Right Side Cover Installation Procedure

Table of Contents

Other manuals for GE Vivid S60N

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid S60N and is the answer not in the manual?

GE Vivid S60N Specifications

General IconGeneral
BrandGE
ModelVivid S60N
CategoryMedical Equipment
LanguageEnglish

Related product manuals