IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
Chapter 3 - System Setup 3-3
PRELIMINARY
3-2-2-1 System Acclimation Time
Following transport, the Vivid S60N/Vivid S70N system may be very cold, or hot. Allow time for the
system to acclimate before being switched ON. Acclimation requires 1 hour for each 2.5
o
C increment,
when the temperature of the system is below 10
o
C or above 40
o
C.
NOTE: After a long period of storage, or after transportation of the system with the monitor in the folded-down
position, it is highly recommended to place the monitor in the upright position - and to leave it in this
position for a period of longer than 1 hour before use. This will enable it to properly adjust to the
environmental conditions.
CAUTION
Turning the system ON after arrival at the site - without allowing time for acclimation - may
cause system damage!
Table 3-7 Vivid S60N/Vivid S70N System Acclimation Time
°C
0 2.5 5 7.5 10 35 40 42.5 45 47.5 50 52.5 55 57.5 60
°F
32 36.5 41 45.5 50 95 104 108.5 113 117.5 122 126.5 131 135.5 140
Hrs
4321002 3 4 5 6 7 8 910

Table of Contents

Other manuals for GE Vivid S60N

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid S60N and is the answer not in the manual?

GE Vivid S60N Specifications

General IconGeneral
BrandGE
ModelVivid S60N
CategoryMedical Equipment
LanguageEnglish

Related product manuals