8-5-21 AC Distribution Box Replacement Procedure
8-5-21-1 Tools
Flat and Phillips screwdriver.
8-5-21-2 Time Required
45 mins
8-5-21-3 Preparation
Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in Power Shut Down on page 4 - 7.
8-5-21-4 AC Distribution Box Removal Procedure
NOTE: The AC Distribution Box is located at the rear of the system - Figure 8-167.
1.) Remove the AC Distribution Box cover, then the battery:
2.) Disconnect the Ground cable - Figure 8-168.
Refer to Table 9-13 on page 9-12.
Figure 8-167 AC Distribution Box Located at Rear of System
• AC Distribution Box Cover Removal Procedure on page 8 - 29
• Battery Removal Procedure on page 8 - 145