IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-109
PRELIMINARY
8-5-10 Front End Power Supply Replacement Procedure
NOTE: The Front End Power Supply (FEPS) is also referred to as the HVPS (High Voltage Power Supply).
8-5-10-1 Tools
Appropriate Phillips screwdriver.
8-5-10-2 Time Required
15 minutes
8-5-10-3 Preparation
Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in Power Shut Down on page 4 - 7.
8-5-10-4 Front End Power Supply Removal Procedure
1) Remove the following covers: left side, right side, DVD, front.
2) Open and remove the Front End door assembly and place it on a stable surface:
3) Loosen the 2 retaining screws that secure the FEPS to the Front End door assembly and gently lift
the plate away from the guide pins located on either side of the FEPS as shown in Figure 8-133.
When performing these procedures, take precautions to avoid damage of
electrostatic-sensitive components. Always have the ESD wrist strap
connected either to the DIB chassis or to the GND plug at the rear of the
scanner, and to your hand.
If a battery is present, first remove the battery as it contains stored energy.
Refer to Battery Removal Procedure on page 8 - 145.
Refer to Table 9-16 on page 9-15.
• Air Inlet (Left Side) Cover Removal Procedure on page 8 - 7
• Right Side Cover Removal Procedure on page 8 - 10
• DVD Cover Removal Procedure on page 8 - 11
• Front Cover Removal Procedure on page 8 - 16
• Front End Door Assembly Opening and Removal Procedure on page 8 - 83
Figure 8-133 Removing the FEPS Metal Cover