IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
3-14 Section 3-3 - Receiving and Unpacking the Equipment
PRELIMINARY
Section 3-4
Preparing for Setup
3-4-1 Verifying Customer Order
Compare items received by the customer to that which is listed on the delivery order. Report any items
that are missing, back ordered, or damaged.
3-4-2 Physical Inspection
Verify that the Vivid™ S60/Vivid™ S70 arrived intact (visual inspection).
If the Vivid™ S60/Vivid™ S70 has been damaged, please refer to DAMAGE IN TRANSPORTATION on
page xiii in the beginning of this manual.
3-4-3 Component Inspection
After verifying that all the required parts are included in the shipping crate, inspect the system components
using the checklist supplied below. In addition, ensure that all the labels described in Chapter 1 -
Introduction are present, accurate and in good condition, and enter the serial number printed on the main
label into the system installation details card, as described in Paperwork After Setup on page 3-56.
3-4-3-1 Damage Inspection Checklist
Visually inspect the contents of the shipping carton for damage. If any parts are damaged or missing,
contact an authorized GE Service Representative.
A Damage Inspection Checklist for the Vivid S60N/Vivid S70N portable ultrasound scanner is provided
in Table 3-9 below.
Table 3-9 Damage Inspection Checklist - Vivid S60N/Vivid S70N Systems
Step Item Recommended Procedure
1
Console
Verify that the system is switched OFF and unplugged. Clean the console and
control panel.
2
Control Console
Physically inspect the control console for missing or damaged items. Verify the
proper illumination of all the control panel buttons.
3
Probes
Check all probes for wear and tear on the lens, cable, and connector. Look for bent
or damaged pins on the connector and in the connector socket on the unit. Verify
that the EMI fingers around the probe connector socket housing are intact. Check
the probe locking mechanism and probe switch.
4
LCD Display
Clean the LCD display by gently wiping with a dry, soft, lint-free non-abrasive
folded cloth. Inspect the monitor for scratches and raster burn.
5
Touch Screen
Clean the Touch Screen by gently wiping with a dry, soft, lint-free non-abrasive
folded cloth. Inspect the screen surface for scratches and raster burn.
6
Fans
Turn on the system and verify that the system’s cooling fans and peripheral fans
are operating.
7
Rear Panel
Check the rear panel connectors for bent pins, loose connections and loose or
missing hardware. Screw all the cable connectors tightly to the connector sockets
on the panel. Verify that the labeling is in good condition.
8
Covers
Check that all screws are tightly secured in place, that there are no dents or
scratches and that no internal parts are exposed.
10
Peripherals
Check and clean the peripherals in accordance with the manufacturer’s directions.
To prevent EMI or system overheating, dress the peripheral cables inside the
peripheral cover.

Table of Contents

Other manuals for GE Vivid S60N

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid S60N and is the answer not in the manual?

GE Vivid S60N Specifications

General IconGeneral
BrandGE
ModelVivid S60N
CategoryMedical Equipment
LanguageEnglish

Related product manuals