8-5-17 Fan (TRX32 Box) Replacement Procedure
8-5-17-1 Tools
Phillips screwdriver.
8-5-17-2 Time Required
15 minutes
8-5-17-3 Preparation
Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in Power Shut Down on page 4 - 7.
8-5-17-4 Fan (TRX32 Box) Removal Procedure
1) Remove the following covers: left side, right side, DVD, front.
2) Open and remove the Front End door assembly and place it on a stable surface:
3) Loosen the 2 screws that secure metal cover on the TRX32 Box and open the cover, as shown in
Figure 8-153 on page 8-127.
When performing these procedures, take precautions to avoid damage of
electrostatic-sensitive components. Always have the ESD wrist strap
connected either to the DIB chassis or to the GND plug at the rear of the
scanner, and to your hand.
If a battery is present, first remove the battery as it contains stored energy.
Refer to Battery Removal Procedure on page 8 - 145.
Refer to Table 9-17 on page 9-16.
• Air Inlet (Left Side) Cover Removal Procedure on page 8 - 7
• Right Side Cover Removal Procedure on page 8 - 10
• DVD Cover Removal Procedure on page 8 - 11
• Front Cover Removal Procedure on page 8 - 16
• Front End Door Assembly Opening and Removal Procedure on page 8 - 83