IRECTION FR091521, REVISION 1VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
Chapter 4 - General Procedures and Functional Checks 4-41
PRELIMINARY
4-3-14-3 Monitor Movement Check
4-3-14-4 Front Wheel Function Check
Table 4-7 Monitor Movement Check
Step Task Expected Result(s)
1
Tilt the monitor forwards and backwards.
Ensure that you do apply some force to move the
monitor. During movement some friction should felt.
During the movement, make sure the image displayed
on the monitor does not present any disturbance.
2
Tilt the monitor sideways - left and right.
Ensure that you do apply some force to move the
monitor. During movement some friction should felt.
During the movement, make sure the image displayed
on the monitor does not present any disturbance.
Table 4-8 Brakes Function Check (Front Castor Wheels)
Step Task Expected Result(s)
1
Release the wheel lock (upper lever) on each front
castor wheel by pushing the lever labeled OFF. Push
and pull the unit right, left, backwards and forwards.
Ensure that the wheels move freely in all directions.
Check the wheels for wear and tear, and replace if
necessary.
2
Press the foot brake (lower lever) down on each front
castor wheel to lock the wheels in position. Push and pull
the unit right, left, backwards and forwards.
Ensure that the wheels are locked and there is no
movement in any direction.

Table of Contents

Other manuals for GE Vivid S60N

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid S60N and is the answer not in the manual?

GE Vivid S60N Specifications

General IconGeneral
BrandGE
ModelVivid S60N
CategoryMedical Equipment
LanguageEnglish

Related product manuals