IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
10-4 Section 10-3 - Why Perform Maintenance Procedures?
PRELIMINARY
Table 10-2 Customer Care & Maintenance Task Schedule
Service at Indicated Time Daily Weekly Monthly
Per
Facility’s
QA Program
Notes
Clean Probes •* * or before each use
Clean Probe Holders
•
Clean Air Filters •
More frequently if necessary, depending on
the environment
Inspect AC Mains Cable
• Mobile Ultrasound system: Check Weekly
Inspect Cables and Connectors
•
Clean Console •
Clean Monitor and Touch Panel •
Inspect Wheels, Casters, Brakes and
Swivel Locks
• Mobile Unit: Check Daily
Check Operator Panel Movement
• Mobile Unit: Check Daily
Console Current Leakage Checks
• Also after corrective maintenance.
Peripheral Current Leakage Checks
• Also after corrective maintenance.
Surface Probe Current Leakage Checks
• Also after corrective maintenance.
Endocavity Probe Current Leakage Checks
• Also after corrective maintenance.
Transesphongeal (TEE) Probe
Current Leakage Checks
• Also after corrective maintenance.
Surgical Probe Current Leakage Checks
• Also after corrective maintenance.
Functional Checks
• Also after corrective maintenance.