EasyManuals Logo
Home>Keithley>Multimeter>3700 Series

Keithley 3700 Series User Manual

Keithley 3700 Series
936 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #136 background imageLoading...
Page #136 background image
Series 3700A System Switch/Multimeter Reference Manual Section 4: Basic DMM operation
3700AS-901-01 Rev. D/June 2018 4-25
Thermocouple connections
Connections for thermocouples are shown below. Thermocouples are color coded to identify the
positive (+) and negative (–) leads (see the table). Note that the negative (–) lead for U.S. type T/Cs is
red.
T/C type
Positive (+)
Negative (–)
T/C type
Positive (+)
Negative (–)
J
White
Red
E
U.S.
Purple
Red
Yellow
Blue
British
Brown
Blue
Red
Blue
DIN
Red
Black
Red
White
Japanese
Red
White
Yellow
Black
French
Yellow
Blue
K
Yellow
Red
R
U.S.
Black
Red
Brown
Blue
British
White
Blue
Red
Green
DIN
Red
White
Red
White
Japanese
Red
White
Yellow
Purple
French
Yellow
Green
N
Orange
Red
S
U.S.
Black
Red
—
—
British
White
Blue
—
—
DIN
Red
White
—
—
Japanese
Red
White
—
—
French
Yellow
Green
T
Blue
Red
B
U.S.
Gray
Red
White
Blue
British
—
—
Red
Brown
DIN
Red
Gray
Red
White
Japanese
Red
Gray
Yellow
Blue
French
—
—
When using the Series 3700A analog backplane connector, use a simulated reference junction for
thermocouple temperature measurements. An ice bath, as shown below, serves as an excellent cold
junction because it is relatively easy to hold the temperature to 0 °C. Notice that copper wires are
used to connect the thermocouple to the Series 3700A input.
Figure 53: Simulated reference junction

Table of Contents

Other manuals for Keithley 3700 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3700 Series and is the answer not in the manual?

Keithley 3700 Series Specifications

General IconGeneral
BrandKeithley
Model3700 Series
CategoryMultimeter
LanguageEnglish

Related product manuals